Pfeiffer HiPace 1200 Turbo Pump with TC 1200 Drive Unit ISO-F 200 Inlet Flange 1300 l/s Pumping Speed Water-Cooled.
Pfeiffer Part Number: PM P06 001.
Pfeiffer HiPace 1200 turbomolecular high-vacuum pumps with an ISO-F 200 inlet flange have a pumping speed of 1300 liters per second (l/s), and an ultimate pressure of 7.5 x 10
-8 Torr. This Pfeiffer HiPace 1200 pump comes with an integrated TC 1200 drive unit mounted underneath the pump and has an ISO KF-40 exhaust foreline flange. The pump can be mounted from vertical (high-vacuum flange on top) to 90° (horizontal). Input power of 100 - 120/200 - 240 VAC 50/60 at a maximum of 1300 Watts is supplied directly to the TC1200 drive unit which powers the pump. The required AC power cable (not included) is part number P1012627. This HiPace 1200 turbo pump with ISO-F 200 inlet and TC 1200 drive unit comes standard with a water-cooling kit and has Pfeiffer part number: PM P06 001.
The included TC 1200 drive unit is a below-pump mounted onboard controller and power supply which includes an RS485 communication interface, a D-sub connector for remote operation, and two accessory ports (to control either a solenoid vent valve or other equipment). For Original Equipment Manufacturer (OEM) applications and system integration, all that is needed to start the pump is the AC input power cable along with start/stop logic. For laboratory use we recommend the Pfeiffer OmniControl 001 turbo pump controller which acts as a remote user interface. The OmniControl 001 can be bench top or rack mounted. It is powered by the TC1200 drive unit through a special RS485 cable, part number P1012287 (sold separately). The OmniControl 001 communicates with and provides full monitoring and control of the HiPace 1200 turbo pump.
We offer a variety of optional accessories, including the OmniControl 001 pump controller, inlet screens, power and communication cables, solenoid operated vent valves, and heater bands (for Conflat CF flanged pumps only).
Pfeiffer HiPace 1200 Turbo Pump with ISO-F 200 inlet and TC 1200 Drive Unit Specifications:
- Part Number: PM P06 001
- Inlet Flange: ISO-F 200 (flanges bolt together)
- Base Pressure: 7.5 x 10-8 Torr
- Operating Speed: 37,800 RPM
- Mounting orientation: 0° (HV flange on top) to 90° (horizontal)
- Pumping Speed: 1300 l/s
- Cooling Method Standard: Water (100 l/h, 15 to 35 °C)
- Water Connection: G 1/4 (BSPP)
- Exhaust Flange: KF-40
- Venting Connection: G 1/8 (BSPP)
- Integral Controller: TC 1200 Drive Unit
- Operating Voltage: 100 - 120/200 - 240 VAC, 50/60 Hz
- Maximum Power: 1350 Watts
- Run-up Time: 2.5 Min.
- Interface: RS485 through M12 Connector
- Remote: High Density 26-Pin Sub-D Connector
- Accessory Ports: Two M12 Sockets, Labeled A and B
- Recommended Roughing Pump: HiScroll 18 Dry Scroll Pump
The Pfeiffer Vacuum HiPace series of turbopumps feature a unique rotor design with proven hybrid bearings. HiPace pumps achieve high pumping speeds and high gas throughput as well as high compression for light gases. HiPace pumps have both turbomolecular and drag stages for improved tolerance to higher backing line pressures and offer enhanced pumping speeds across all gas species. HiPace hybrid bearings are a combination of a ceramic ball bearing on the fore vacuum side and a permanent magnetic radial bearing on the high-vacuum side. This bearing technology does not require electromagnets and has an extended service life with typical maintenance intervals of approximately 4 years. In addition to enhanced performance, all HiPace turbo pumps have field serviceable oil cartridges and shaft bearings which dramatically decreases downtime.
Pfeiffer HiPace pumps with a Conflat CF flange are true ultra-high vacuum (UHV) pumps with an ultimate pressure into the 10
-10 Torr range (if all-metal intake seal gaskets are used and the vacuum chamber is completely cleaned and baked out). Turbo pumps with ISO flanges are limited (by permeation and outgassing of the rubber/elastomer O-ring seals) to an ultimate pressure in the 10
-8 Torr range. The intake of Conflat-flanged HiPace pumps can be baked up to 120 °C if the turbo is water cooled.
Typical Turbo Pump Applications include:
- General clean pumping applications
- Scanning Electron Microscopes - SEM
- Beam lines and high energy physics
- Mass Spectrometry
- Electron Microscopy
- Sample Preparation
- Research and Development
- High Energy Physics
- Industrial
- Semiconductor
- Coatings
- Surface Science Instruments
The pump instruction manual and Pfeiffer HiPace pump product brochures are available in the downloads.