Skip to main content
× Home Products Downloads About Contact Careers (505) 872-0037

(505) 872-0037

Shopping Cart Icon
cart
 
Login Icon
login

Keyword       Ideal Spectroscopy Part Number       Manufacturer Part Number

× Molecular-Spectroscopy Optical-Table-Accessories Absorption-Spectroscopy-Gas-Cells Ultra-High Purity Gas Products Pulse-Valves-For-Jet-Cooling Vacuum-Pressure-Measurement Fittings-and-Flanges Vacuum-Valves Vacuum-Pumps Modular-Vacuum-Chambers

Condition:
  New
Part Number:
  P1012514
Warranty:
  Edwards Warranty
Out of Stock  



$1,176.82


Print Version

Edwards 240 VAC Flange Heater Band for nEXT Turbo Pumps with CF 4.5 in. DN 63 CF Inlet Flange. PN: B8G200824


Edwards Flange Heater Band, CF 4.5", DN 63 CF, 240 VAC, for nEXT85D Turbo Pumps.
Edwards Part Number: B8G200824.

This is a 240 VAC Edwards intake flange heater for the nEXT85D turbo pump with CF 4.5 inch flange (DN 63 CF). This heater band may only be used on CF flanged pumps outfitted with either water cooling or forced air cooling. The intake flange may be baked up to 120 °C if the turbo pump is water cooled, 115 °C if the turbo pump is fan cooled. This flange heater (bakeout band) accelerates the degassing of the pump to enable it to achieve lower pressures. It may also be used to protect the pump from condensation of contaminants. The band heater has an attached power cable terminated with a C19 power entry plug (male C19). This Edwards intake bakeout flange heater band has part number B8G200824.

We recommend using the Ideal Vacuum PID plug and play heater controller (part number P107361). The heater band power cable will plug directly into it. Our P107361 controller can be operated at 100-240 VAC. The input voltage of the controller must be the same as the heater. For use with 240 volt applications you must supply your own power cord.

Heater Band Specs:
  • Fits: nEXT85D turbo pump with CF 4.5 in. flange (P109212) and cooling option (required)
  • Water Cooling Option: P1012512
  • Forced Air Cooling Option: P109582
  • Heater Band Voltage: 240 VAC
  • Recommended Controller: Ideal Vacuum P107361 (requires customer supplied power cable)



Press ESC to Close


PDF Logo Edwards nEXT 55D 85D Turbo Pumps Instruction Manual.pdf

PDF Logo Edwards nEXT 85D Turbo Vacuum Pump Datasheet.pdf

PDF Logo Edwards nEXT Turbomolecular Pumps Entire Range Brochure.pdf

PDF Logo Edwards TIC Relay Box Instruction Manual.pdf

PDF Logo Edwards TAG Turbo and Active Gauge Controller Instruction Manual.pdf

PDF Logo Edwards TIC Turbo Instrument Controller Instruction Manual.pdf



You may wish to consider the following products:

Ideal Vacuum Logo Water Mark
CONTACT US
Ideal Spectroscopy
5910 Midway Park Blvd NE,
Albuquerque, NM 87109-5805

Phone: (505) 872-0037
Fax: (505) 872-9001
info@idealspectroscopy.com



Press ESC to Close


deepbox image 2