This is a 240 VAC Edwards intake flange heater for the nEXT85D turbo pump with CF 4.5 inch flange (DN 63 CF). This heater band may only be used on CF flanged pumps outfitted with either water cooling or forced air cooling. The intake flange may be baked up to 120 °C if the turbo pump is water cooled, 115 °C if the turbo pump is fan cooled. This flange heater (bakeout band) accelerates the degassing of the pump to enable it to achieve lower pressures. It may also be used to protect the pump from condensation of contaminants. The band heater has an attached power cable terminated with a C19 power entry plug (male C19). This Edwards intake bakeout flange heater band has part number B8G200824.
We recommend using the Ideal Vacuum PID plug and play heater controller (part number P107361). The heater band power cable will plug directly into it. Our P107361 controller can be operated at 100-240 VAC. The input voltage of the controller must be the same as the heater. For use with 240 volt applications you must supply your own power cord.
Heater Band Specs:
- Fits: nEXT85D turbo pump with CF 4.5 in. flange (P109212) and cooling option (required)
- Water Cooling Option: P1012512
- Forced Air Cooling Option: P109582
- Heater Band Voltage: 240 VAC
- Recommended Controller: Ideal Vacuum P107361 (requires customer supplied power cable)