Skip to main content
× Home Products Downloads About Contact Careers (505) 872-0037

(505) 872-0037

Shopping Cart Icon
cart
 
Login Icon
login

Keyword       Ideal Spectroscopy Part Number       Manufacturer Part Number

× Molecular-Spectroscopy Optical-Table-Accessories Absorption-Spectroscopy-Gas-Cells Ultra-High Purity Gas Products Pulse-Valves-For-Jet-Cooling Vacuum-Pressure-Measurement Fittings-and-Flanges Vacuum-Valves Vacuum-Pumps Modular-Vacuum-Chambers

Condition:
  New
Part Number:
  P103389
Warranty:
  Pfeiffer Warranty

Available Now: 
2  



$9,517.40


Regular Price: $11,330.24

On Sale Flyer
Print Version

Pfeiffer HiPace 80 Turbo Pump DN 63 ISO-K 63 Inlet Package Deal with TC 110 Drive Unit OmniControl 200 and Air Cooling Kit


Pfeiffer HiPace 80 Turbo Pump DN 63 ISO-K 63 Inlet Package Deal with TC 110 Drive Unit OmniControl 200 and Air Cooling Kit.

This Pfeiffer HiPace 80 turbo pump package deal with DN 63 ISO-K 63 inlet flange has a pumping speed of 67 l/s and an ultimate pressure of 7.5 x 10-8 Torr). The pump has an integrated TC 110 electronic pump drive unit. This package includes an OmniControl 200 display control unit and power supply with large LED display (includes AC power cable). The OmniControl 200 powers, communicates with, and controls the pump through an included 3 meter long controller connection cable (part number P103335) with incorporated M8 ports for accessories and an M12 connector for connecting the included RS485 communications cable (part number P1012287). This deal also includes a radial air cooling fan unit which plugs into one of the M8 ports on the controller connection cable and is powered by the pump. To protect the pump from debris, a splinter guard with fine mesh (0.8 mm) screen is also part of this package. This turbo pump package deal does not include a roughing pump, roughing lines, vacuum chamber, pressure measurement gauges, or valves. Some upgrades might include: a water cooling kit, solenoid venting or sealing gas purge valves. All these items are available and sold separately by Ideal Vacuum.

About the HiPace 80 turbo pump with DN63 ISO-K flange
Pfeiffer HiPace 80 turbomolecular high-vacuum pumps with a DN63 ISO-K inlet flange and KF-16 outlet flange have a pumping speed of 67 liters per second (l/s) and an ultimate pressure of 7.5 x 10-8 Torr. The pump can be mounted in any orientation. 24 VDC input power at a maximum of 110 Watts is required. This Pfeiffer HiPace 80 pump comes with an integrated TC 110 drive unit side-mounted to the pump, which has a single X3 (15-pin D-Sub) connector for supplying pump power, RS485 communications and M8 accessory ports (to control either an air-cooling fan or solenoid vent valve). This HiPace 80 turbo pump with ISO-K 63 inlet and TC 110 drive unit has Pfeiffer part number: PM P03 940 A.

About the OmniControl 200 drive control unit
The Pfeiffer OmniControl 200 turbo pump controller can be bench top or rack mounted and is designed to operate Pfeiffer HiPace 80 and HiPace 300H turbo pumps that have the integrated TC110 drive unit. The OmniControl 200 supplies 24 VDC at a maximum of 200 Watts to the TC 110 electronic drive unit which then powers the pump. The required connection cable between the OmniControl 200 and TC 110 is included. The connection cable not only supplies power to the pump it also has M8 ports for connecting the included radial forced air cooling fan unit, and/or a sealing gas valve. Input power to the OmniControl 200 is 115/230 VAC and a standard 115 VAC power cable is included. This OmniControl 200 turbo pump display control unit and power supply has Pfeiffer part number PE D50 100 0.

About the radial air-cooling fan unit
This Pfeiffer air cooling fan unit mounts to the side of HiPace 80 turbo pumps and provides air flow perpendicular to the pump. This air cooling fan plugs into one of the accessory ports on the TC 110-to-OmniControl cable. The air cooling unit always runs when the pump is on. This radial air cooling fan unit has Pfeiffer part number PM Z01 300 AT.

Applications for this package deal include:
  • Mass Spectrometry
  • Gas Chromatography Mass Spectrometry (GC/MS)
  • Liquid Chromatography Mass Spectrometry (LC/MS)
  • Inductively Coupled Plasma Mass Spectrometry (ICP-MS)
  • Time-of-Flight Mass Spectrometry (TOF)
Vacuum in Research and Development:
  • High Energy Physics
  • Fusion Technology
  • General UHV Research
  • Synchrotron Light Sources Particle Accelerators
Nanotechnology Instruments:
  • Electron Microscopy
  • Scanning Electron Microscopy (SEM)
  • Transmission Electron Microscopy (TEM)
  • Focused Ion-beam Systems (FIB)
  • Surface Analysis
  • Semiconductor Manufacturing
Industrial Vacuum Processing:
  • Thin Film Deposition:
  • Glass coating equipment (architectural and automotive glass, flat panel displays)
  • Thin film solar cells production (photo-voltaic)
  • Optical data media (CD, DVD, Magneto Optical Discs)
  • Magnetic storage media
  • Surface treatments
  • Optical coating (ophthalmic, precision opto-electronics)
  • Roll/Web coating on films or foils
Device Processing:
  • TV & Monitor picture tube manufacture
  • Evacuation of lamps (motorway lighting, beamers)
  • X-Ray tubes & electron devices
General Processes:
  • Vacuum Furnaces
  • Metallurgy
This package deal is stored in our inventory as a kit, made up of many separate parts, and each is listed in our Availability of Kit Components section below. The operating instructions manuals are available in the downloads.



Press ESC to Close




Availability Of Kit Components

Kit: P103389 - Pfeiffer HiPace 80 Turbo Pump DN 63 ISO-K 63 Inlet Package Deal with TC 110 Drive Unit OmniControl 200 and Air Cooling Kit
Component
Desc.
Req'd
Avail
RS485 Communication Cable M12 to M12 Connectors, Connects Pfeiffer HiPace Turbo Pump TC Drive Unit to OmniControl Controller, 3 m Long.
1
39
Pfeiffer OmniControl 200 Turbo Pump Display Control Unit and Power Supply 220W 24 VDC. PN: PE D50 100 0
1
2
Pfeiffer HiPace 80 Turbo Pump with TC 110 Drive Unit DN63 ISO-K 63 Inlet Flange 67 l/s Pumping Speed Air-or Water-Cooled, PN: PM P03 940 A
1
23
Pfeiffer Radial Air Cooling Fan Unit for HiPace 80 Turbo Pumps with TC 110 Drive Unit Powered by M8 Accessory Connector, PN: PM Z01 300 AT
1
114
Pfeiffer TC 110 Electronic Turbo Pump Drive Unit to OmniControl 200 Controller Connection Cable with Two M8 Accessory Ports 3 m Long, PN: PM 061 351-T
1
21
Pfeiffer Fine Inlet Centering Ring Screen (0.8 mm) Splinter Guard fits HiPace Turbo Pumps with DN 63 ISO-K 63 Flange, PN: PM 016 207 AU
1
6
Pfeiffer Turbo OmniControl Controller and HiScroll Roughing Pump Power Supply Cord Mains Cable 115 VAC NEMA 5-15 to IEC C13, 3 m Long, PN: P 4564 309 ZE
1
98
PDF Logo Pfeiffer HiPace 80 Turbopump with TC 110 Electronic Drive Unit Operating Instructions.pdf

PDF Logo Pfeiffer TC 110 Turbopump Electronics Drive Unit Operating Instructions.pdf

PDF Logo Pfeiffer OmniControl Turbo Pump Controller Operating Instructions.pdf



You may wish to consider the following products:

Ideal Vacuum Logo Water Mark
CONTACT US
Ideal Spectroscopy
5910 Midway Park Blvd NE,
Albuquerque, NM 87109-5805

Phone: (505) 872-0037
Fax: (505) 872-9001
info@idealspectroscopy.com



Press ESC to Close


deepbox image 2