Agilent Varian HS-16 High Vacuum Diffusion Pump with Halo Baffle, 10,000 l/s Pumping Speed
16 Inch ASA Inlet, 3 Inch ASA Exhaust, 240 VAC
These Varian Agilent High Throughput models include the series HS-16, HS-20, HS-32 and NHS-35. This is the HS-16 diffusion high throughput vacuum pump. They are the primary industrial standard for high vacuum pumping mechanism. The Agilent Varian HS-16 diffusion pumps have a high pumping speed for all gases and low cost per unit pumping speed when compared with other types of pump used in the same vacuum range. The full fractionating design allows vaporized fluid to be fractionated in the boiler and jet assembly. Contamination and decomposition products are pumped away and only the purest vapor reaches the top jet, assuring highest speed and lowest backstreaming. The vacuum fluid heater is mounted at the bottom of the pump body. The HS pumps have a fill and drain assembly for easy maintenance of pump fluid and continuous monitoring with thermal protection switches. Finned boilers increase the surface area which prolongs fluid life. The inlet is at the top, and the exhaust is through the foreline.
Diffusion pumps use a high-speed jet of vapor to direct gas molecules in the pump throat down into the bottom of the pump and out the exhaust. The oil diffusion pump is operated with an oil that has low vapor pressure. Its purpose is to achieve higher vacuum (lower pressure) than possible by use of positive displacement pumps alone. Although its use has been mainly associated within the high vacuum range (down to 10-9 Torr), diffusion pumps today can produce pressures approaching 10-10 Torr when properly used with modern fluids and accessories. The features that make the diffusion pump attractive for high and ultra-high vacuum use are its high pumping speed for all gases and low cost per unit pumping speed when compared with other types of pump used in the same vacuum pressure range. Diffusion pumps cannot discharge directly into the atmosphere, so a mechanical forepump is typically used to maintain an outlet pressure around 1 mTorr.
These Varian model HS-16 high vacuum, high throughput diffusion pumps are new and can be ordered with either a 3-phase 240, 415 or 480 VAC heaters. When ordering please specify which voltage your application will require. We stock and sell diffusion pumps oils, e.g. IVP 704 diffusion pump fluids along with vacuum fittings, adapters, and replacement parts for your Agilent Varian HS-16 diffusion pumps. This pump uses IVP 704 diffusion pump fluid and is designed for high vacuum and fast pumping of large volumes of gas or vapor in production operations.
For HS-16 Instruction Manual scroll down to manuals tab AVAILABLE DOWNLOADS and download the PDF.
Highlights For This Agilent Varian HS-16 Diffusion Pump
16 Inch ASA Inlet Flange, 23.5 in. OD, 1.12 in. hole size (16 bolt holes)
ASA 3 in Outlet Flange, 7.5 in. OD, .75 in. hole size (4 bolt holes)
Recommended Roughing Backing Pump: ≥80 CFM
Halo Baffle
Maximum Pumping Speed 10,000 l/s air
240V @ 2700 Watt Heater (3 required)
Designed To Be Used With IVP 704 Diffusion Pump Oil (3 Quarts)
Included with ASA Pumps:
One Inlet Flange O-ring PN 48214001
One Foreline Flange O-ring PN 660890348
Two 1/2" Hose 3/8 Pipe Barbed Connectors PN 624301111
Four 3/8" Hose 3/8 Pipe Barbed Connectors PN 624301110
Installation and Operation Manual PN 699901140
Press ESC to Close
ITEM
DESCRIPTION
PART NO.
1
Standard Cold Cap Assembly - Includes 10-32 SST Rd Hd screw & 10-32 SST Hex nut
79299301
1
Water-Cooled Halo Baffle
K0143316
2
Cold Cap Water Connection:
Cold Cap Gasket/Grommet (Kit of 4)
L8839301
Cold Cap Nut & Followers Kit
L8840301
3
Jet Assembly
83612401
4
Jet Coupling
L6367301
5
Foreline Baffle Assembly
80798301
6
Sight Glass Service Kit for Pumps Built Before 1995:
3 Glass Tubes
F1755008
2 Brass Union 1/4 OD x 1/4 MPT
622568108
2 Brass Elbow 69FL 6x4, 169VL6-4
623451012
8 O-ring, Viton, Parker No. 2-011
660892011
8 O-ring, Viton, Parker No. 2-111
660892111
3 O-ring, Viton, Parker No. 2-221
660892221
3 O-ring, Viton, Parker No. 2-213
660892213
F6097301
6
Sight Glass Service Kit for Pumps Built After 1995
Ideal Spectroscopy offers 3D CAD models in Solidworks (SLDPRT), STP, & STEP formats. For those without CAD software a free trial 3D model viewer with measurement capabilities can be
downloaded here: