This is a Matheson Single Stage Chrome Plated Brass regulator with a 1/4 in compression tube outlet connection and 1/4 in FNPT body porting. It comes equipped with 2-1/2 inch inlet and delivery pressure gauges, an outlet diaphragm valve, metal to metal seals, chrome plated bonnet, and a Teflon seat.
This Matheson series 3530A, model 3536A regulator is to be used in applications requiring delivery pressures up to 250 PSIG and can accept a maximum inlet pressure of 3000 PSIG. This regulator is suitable for a variety chromatorgraphy applications requiring the supply of carrier gas/detector support gas. This regulator is not suitable for any application requiring the supply of corrosive gas (non-corrosive gas applications only).
Gas Chromatography Detectors which can use the 3536A Regulator:
- Flame Ionization Detector (FID)
- Thermal Conductivity Detector (TCD)
- Nitrogen Phosphorus Detector (NPD)
- Flame Photometric Detector (FPD)
- Photoionization Detector (PID)
- Electrolytic Conductivity Detector (ELCD or Hall Detector)
- Electron Capture Detector (ECD)
Single-Stage vs. Two-StageBoth single-stage and two-stage regulators operate by reducing the pressure of gas supplied by a cylinder to an appropriate output level. The intake and delivery pressure ranges of regulators do not depend on how many stages they have, but by each regulators mechanical characteristics. When considering a single vs. a two-stage regulator there are two things to keep in mind:
Droop and
Supply Pressure Effect. Droop is the decrease in output pressure caused by an increase in flow rate. Supply Pressure Effect is simply the variation in delivery pressure from the cylinder to the regulator.
Single Stage regulators will experience less droop with varying flow rates but a large supply pressure effect and need to be monitored and adjusted by an operator or used in applications where supply pressure will remain constant. Two-Stage regulators inherently control for variations in supply pressure while producing a constant output and therefore can be used without an operator present and in applications which require a very specific delivery pressure. The specification sheet for this series of regulator can be found in the downloadable section as a .PDF.