Agilent Varian IDP7, IDP10, IDP-07/IDP-10 Dry Scroll Vacuum Pump Tip Seal Replacement Kit. Part Number X3807-67000
These rebuild kits are used during maintenance on Agilent IDP-07 and IDP-10 dry scroll vacuum pumps. The Agilent IDP-07 and IDP-10 pumps will provide many years of trouble-free service if the maintenance procedures and intervals are observed. Cleaning and tip seal replacement are recommended when pump base pressure has risen to an unacceptably high level for your application. If your pump exhibits humming or grinding noises from the bearings, a major overhaul should be done by Agilent personnel or an authorized rebuild center. The parts needed for tip seal replacement on the IDP-07 and IDP-10 are available in this kit as described; replacement tip seals, Viton O-rings, assorted hex cap screws and gas ballast fittings. Instruction manuals and brochures in PDF format for the new Agilent IDP-07 and IDP-10 hermetically-sealed dry scroll vacuum pumps can be downloaded in our available download section below.
The tip seal replacement for both IDP-07 and IDP-10 pumps is easier than ever before. All that is needed is these simple tools, flat head screw driver, 4 and 5 mm Allen wrenches, isopropyl alcohol, lint-free wipes, and this tip seal kit part number X3807-67000.
|