This is a 115 VAC Leybold intake flange heater for the HiPace 300 and 350 turbo pump (including chemical resistant models) with CF 6.0 inch flange (DN 100 CF) and TC 400 drive unit. The intake flange may be baked up to 120 °C if the turbo pump is water cooled. Operate the pump with water cooling when baking out. Do not use this intake flange heater band on an air cooled pump - damage to the pump will occur.
The band heater has an attached relay box with a short control cable with M12 connector which connects directly to one of the accessory ports on the TC 400 drive unit. The relay box must be wired with appropriate cable and plug to a 115 VAC power source.
This heater band for the HiPace 300 and 350 turbo pump with Conflat CF 6.0" flange has part number PM 061 368-T.
Heater Band Specs:
- Fits: HiPace 300 turbo pump with CF 6.0 in.flange and TC 400 drive unit (P103387 or P105445)
- Fits: HiPace 350 turbo pump with CF 6.0 in.flange and TC 400 drive unit (P1012600)
- Water Cooling Option (required): P103364 (included with pump)
- Heater Band Voltage: 115 VAC
- Accessory Connection: M12
- Recommended Controller: OmniControl 300